Operating Manual

Accu-Wave Chapter 3 Set up Level, Level Alarms, and Volume
TN Technologies 3-3
“Set up Level” menu. The “Comments” column of the table describes the purpose of each
menu item. An example showing the specific entries required to set up level as the primary
measurement is provided immediately following this table.
“Set up Level” Menu
Display Comments
Set up level, level
alarms & volume
Exit Setup.
Other functions
From the measurement display, press to move to
the “Set up level,...” menu shown at the left. Press
to access the “Set up level,…” menu items.
Note: The software detects whether output relays
are installed. If no relays are present, the
menu will display “Set up level & volume.”
General HELP text.
Information on how
to set up this gauge
NEXT
We strongly suggest you read the HELP screens,
especially when you are first learning to set up the
gauge. Press to proceed to the next menu item.
primary measurement:
Distance
Change to Level
NEXT
The primary measurement can be either distance or
level. Distance is the default.
Allow display of ALL
units. Change to
Metric or English
NEXT
Select the desired set of units for the primary
measurement using the key. The default is ALL
which includes both English and Metric units.
English units are - in, ft, and yd. Metric units are -
mm, cm, and M.
Distance units = ft
To change to yd, M,
cm, mm, or in press
NEXT
Scroll through and select desired measurement units
using the key. The available units depend on
previous menu item selection of ALL, English, or
Metric units.
Note: The next two menu items define the Range of Interest, the range over which
the gauge looks for the echo from the process material surface. The values
to be entered depend on whether distance or level is selected as the primary
measurement.