Operating Manual
Accu-Wave Chapter 9 Security, Service & Diagnostic
TN Technologies 9-9
User Service & Related Items
“User Service & Related Items” Menu
Display Comments
System restart. Does
not affect user data
←←Exit this menu
NEXT↓↓ EXECUTE CMD→→
Re-initializes system including the DSP. Erases
temporary memory, but does not affect user-entered
setup data.
Erase all entries !!!
(except COMM setup)
←←Exit this menu
NEXT↓↓ EXECUTE CMD→→
Erase all previously entered set up data. All settings
except for serial communication settings are reset to
factory defaults. Gauge will behave just like the first
time power was applied.
Clear all holds
(none now in effect)
←←Exit this menu.
NEXT↓↓ EXECUTE CMD→→
Execute this command to clear all holds (if any holds
are in effect).
Hold current output
at max (20.00 mA)
←←Exit this menu.
NEXT↓↓ EXECUTE CMD→→
Holds all current outputs at the value entered for the
maximum current output.
Hold current output
at min (4.000 mA)
←←Exit this menu.
NEXT↓↓ EXECUTE CMD→→
Holds all current outputs at the value entered for the
minimum current output.
Hold current output
at zero mA
←←Exit this menu.
NEXT↓↓ EXECUTE CMD→→
Holds all current outputs at 0.0 mA.
Current output
hold mode value
50.00% of scale
NEXT↓↓
Enter value in % for the mid-range current output
hold value. Default value is 50.00%.
Hold current output
at 50.00% of scale
←←Exit this menu.
NEXT↓↓ EXECUTE CMD→→
Press → to execute command to hold current output at
the mid-range value specified in the previous menu
item.