Operating Manual
Chapter 6 Cuurent Outputs, Relays, and Alarms Accu-Wave
6-2 TN Technologies
“Modify or Re-assign Current Output” Menu
Display Comments
Modify or re-assign
current output↓↓
Other functions→→
Press ↓ to access “Current Output” menu items. Press
→ to scroll to other menus.
maximum current
output 20.00 mA
(4.000 to 20.00)
NEXT↓↓
The range for the maximum current output is from
the minimum current output (currently 4.000mA) to
20.00mA. The default value for the maximum current
output is 20.00mA.
minimum current
output 4.00 mA
(.0001 to 20.00)
NEXT↓↓
The range for the minimum current output is from
.0001mA to the maximum current output (currently
20.00mA). The default value for the minimum
current output is 4.000 mA.
Note: Entering a value of exactly 0.0 causes the
default value of 4.000mA to be used for the
minimum current output.
Note: If two or more measurements have been set up, the next two menu items
will be displayed. If a second current output is also installed, the next two
menu items will be repeated for “current out 2.”
Mea 1: ft level
is sent to current
out 1 in normal
mode. NEXT↓↓ CHANGE→→
Assign desired measurement to current output 1 in
the normal mode. Select from the primary
measurement and the measurements set up using the
“Additional measurements” menus.
Mea 1: ft level
is sent to current
out 1 in alternate
mode. NEXT↓↓ CHANGE→→
Assign a measurement to current output 1 in the
alternate mode. If system includes more than one
current output, additional entries will show up.
Note: If a second current output is installed, the next two menu items will be
displayed first for “current output 1” and then for “current output 2.”
correction factor
for current output
at maximum: 1.000
NEXT→→
Fine tune the maximum current output value to
correct for component and wiring variations. (The
maximum current output value is scaled by this
value.) This item is repeated for each current output
installed.
correction factor
for current output
at minimum: 1.000
NEXT→→
Fine tune the minimum current output value to
correct for component and wiring variations. (The
minimum current output value is scaled by this
value.) This item is repeated for each current output
installed.
current output
hold mode value
50.00 % of scale
NEXT↓↓
Enter desired value for mid-range hold value for the
current output. Value is entered as a percentage of the
maximum current output value. The default is 50%.