Operating Manual

Chapter 9 Security, Service & Diagnostics Accu-Wave
9-10 TN Technologies
“User Service & Related Items” Menu (cont.)
Display Comments
Level hold mode
value
15.00 ft level
NEXT
Enter the hold value for the primary measurement
(level or distance). This value is also used as the “lost
echo value” when “go to hold value” is selected as the
“echo loss mode.” See page 5-8.
Hold level at
15.00 ft level
Exit this menu.
NEXT EXECUTE CMD
Press to hold primary measurement value at the
hold value specified in the previous menu item. After
executing command, display changes to read “Clear
holds.” Press to clear the hold level command.
level hold value
0.000 cm
{for abs level test}
{cmd 154} NEXT
[Service only item]
Enter hold value for level used to test the absolute
level. Execute level hold using the next menu item.
hold level at
0.000 cm
Exit this menu.
NEXT EXECUTE CMD
[Service only item]
Press to execute command to hold primary
measurement value at the value specified in previous
item. Use “Clear all holds” to clear.
Volume hold mode
value 150.0 cu ft
NEXT
Only displayed if “tank volume” setup has been
completed. Enter the volume hold value.
Hold Volume at
150.0 cu ft
Exit this menu.
NEXT EXECUTE CMD
Press to execute command to hold volume
measurement value at the hold value specified in the
previous menu item.
Review measurement
assignments 1-8
NEXT
Press to view current measurements, e.g.,
Meas #s 1, 2, & 3
Level
volume cu ft
(inactive)
Press to view other measurements if setup.
Do enable
service only items.
Change to “Do”
Continue as is.
Change to “Do not” to disable “service only items.”
This will disable a variety of additional menu items
throughout the Setup menus. In particular, several of
the Radar Fine Tuning tools will be disabled
including “Tank Map Setup,” Region Weighting
Setup,” and the “Dynamic Threshold Setup.”