Operating Manual
Accu-Wave Chapter 9 Security, Service & Diagnostic
TN Technologies 9-7
“Snapshot” Menu (cont.)
[Service only item]
Display Comments
base
signal strength
113.9
NEXT↓↓
Signal Strength of “best” candidate echo as reported
in the echo report.
weighted
signal strengh
113.9
NEXT↓↓
Signal strength of “best” candidate echo after
weighting factors have been applied.
View echos being
processed→→
NEXT↓↓
Allows you to examine the echo report for each echo
within the range of interest. Press → to access the
echo reports, then press ↓ to scroll through the echo
reports. The echo report is described on page 5-11.
View marker report→→
NEXT↓↓
Reports relative control levels at which RF frequency
markers occur. See “View Marker Report” menu table
below.
DSP restart count
0
NEXT↓↓
Internal value of
Sensor signal
1.871E4
←←CONT UPDATE→→ NEXT↓↓
Filtered value of echo position after electrical offset is
subtracted. Press ← for continuous measurement
updates. Press ← again to return to FREEZE mode.
Internal value of
distance
441.0 cm
←←CONT UPDATE→→
Filtered value of echo position (distance) converted to
centimeters. Most internal values in the software are
stored in CGS units (centimeters, grams, seconds).
Note: The next three menu items will only display meaningful values when a
single current output board is installed.
Internal value of
I out %
60.80 %
←←CONT UPDATE→→
Internal value of current output - % of range.
Internal value of
I out flt
5429 (8000=100%)
←←CONT UPDATE→→
Internal value of
I out fix
5428 (8000=100%)
UPDATE→→
This is the value actually sent to the current output
board.