Instructions

The 5.5 +5°C refrigerator has a small drain tube in the rear of the unit that may become blocked by debris and cause
condensation to back up into the rear of the unit. Follow Instructions below for unblocking a clogged drain tube.
The TSX505 & TSG505 refrigerator green residue or water pooling in the bottom interior.
Note: The source of the inorganic green residue material is plating of an internal heat exchanger and a
oxidized nickel . blocked drain tube. It is not harmful to the cooling performance of the product and
clean with a cloth and nonabrasive cleaning solution.