Instruction Manual

User Calibration Installation and Operation Handbook
7-2 2416 Controller
7.2 USER CALIBRATION ENABLE
The User calibration facility must first be enabled in configuration level by setting the
parameter ‘UCAL' in the CAL ConF list to 'YES'. This will make the User calibration
parameters visible in Operator ‘FuLL’ level. This procedure is described in Chapter 6,
Configuration, but for convenience is summarised below: .
+
The Calibration Configuration List
Press until you reach the ‘CAL-Conf’ list.
Press until you reachUCAL’.
User Calibration Enable
Use or to select:
YES: Calibration enable
no: Calibration disabled
Press
and together to go to the Exit display.
Exit configuration
Use
or to select ‘YES’ to return to Operator level.
CAL
UCAL
YES
×2