Datasheet

TLC0831C, TLC0831I
TLC0832C, TLC0832I
8-BIT ANALOG-TO-DIGITAL CONVERTERS WITH SERIAL CONTROL
SLAS107B – JANUARY 1995 – REVISED APRIL 1996
7
POST OFFICE BOX 655303 DALLAS, TEXAS 75265
operating characteristics V
CC
= V
ref
= 5 V, f
clock
= 250 kHz, t
r
= t
f
= 20 ns, T
A
= 25°C (unless otherwise
noted)
PARAMETER
TEST CONDITIONS
MIN TYP MAX UNIT
Supply-voltage variation error V
CC
= 4.75 V to 5.25 V ±1/16 ±1/4 LSB
Total unadjusted error (see Note 5)
V
ref
= 5 V,
T
A
= MIN to MAX
±1 LSB
Common-mode error Differential mode ±1/16 ±1/4 LSB
Propagation delay time,
MSB-first data
C
L
= 100
p
F
650 1500
ns
pd
gy,
output data after CLK(see Note 6)
LSB-first data
C
L
=
100
pF
250 600
ns
Output disable time DO after CS
C
L
= 10 pF, R
L
= 10 k 125 250
ns
dis
O
u
t
pu
t
di
sa
bl
e
ti
me,
DO
a
ft
er
CS
C
L
= 100 pF, R
L
= 2 k 500
ns
t
conv
Conversion time (multiplexer-addressing
time not included)
8
clock
periods
All parameters are measured under open-loop conditions with zero common-mode input voltage. For conditions shown as MIN or MAX, use the
appropriate value specified under recommended operating conditions.
NOTES: 5. Total unadjusted error includes offset, full-scale, linearity, and multiplexer errors.
6. The MSB-first data is output directly from the comparator and, therefore, requires additional delay to allow for comparator response
time. LSB-first data applies only to TLC0832.