Use Instructions

SpiroSphere
®
ECG Instructions for Use
Page 68/136
Version 02.02 • 04AUG2022
The electrodes can be visually checked (“running lights”), starting with “V1” and end with “LL”.
Look for the “e” shape to conrm correct lead placement.
Press <Next>.
Prior to the ECG recording, an Impedance Check will be performed to conrm good contact with
the skin. If a contact is poor, the respective electrode will blink on the screen:
If an electrode contact is indicated as poor, please check the respective contact.