Datasheet

OUTPUTT VOLTTAGE (1V/DIV)
TIME (25ms/DIV)
INPUT PULSE (100mV/DIV)
LMP91000 OUTPUT
TEST PULSE
LMP91000
SNAS506H JANUARY 2011REVISED MARCH 2013
www.ti.com
SENSOR TEST PROCEDURE
The LMP91000 has all the hardware and programmability features to implement some test procedures. The
purpose of the test procedure is to:
a. test proper function of the sensor (status of health)
b. test proper connection of the sensor to the LMP91000
The test procedure is very easy. The variable bias block is user programmable through the digital interface. A
step voltage can be applied by the end user to the positive input of A1. As a consequence a transient current will
start flowing into the sensor (to charge its internal capacitance) and it will be detected by the TIA. If the current
transient is not detected, either a sensor fault or a connection problem is present. The slope and the aspect of
the transient response can also be used to detect sensor aging (for example, a cell that is drying and no longer
efficiently conducts the current). After it is verified that the sensor is working properly, the LMP91000 needs to be
reset to its original configuration. It is not required to observe the full transient in order to contain the testing time.
All the needed information are included in the transient slopes (both edges). Figure 35 shows an example of the
test procedure, a Carbon Monoxide sensor is connected to the LMP91000, two pulses are then sequentially
applied to the bias voltage:
1. from 0mV to 40mV
2. from 40mV to -40mV
and finally the bias is set again at 0mV since this is the normal operation condition for this sensor.
Figure 35. Test Procedure Example
24 Submit Documentation Feedback Copyright © 2011–2013, Texas Instruments Incorporated
Product Folder Links: LMP91000