Typical Application
Figure 28. pH Probe Amplifier/Temperature Compensator
***For R2 = 50k, R4 = 330k ±1%
For R2 = 100k, R4 = 75k ±1%
For R2 = 200k, R4 = 56k ±1%
**Polystyrene
*Film resistor type RN60C
To calibrate, insert probe in pH =7 solution. Set the “TEMPERATURE ADJUST” pot, R2, to correspond to the solution
temperature: full clockwise for 0°C, and proportionately for intermediate temperatures, using a turns-counting dial.
Then set “CALIBRATE” pot so output reads 7V.
Typical probe = Ingold Electrodes #465-35
Detailed Schematic
Figure 29. 1/4 Quad