Pro (Advanced) Screen
9.4 Pack Voltage Calibration
To calibrate the voltage at the AFE Pack pin, follow these steps:
1. Ensure that Voltage Calibration has been performed for the pack. Ensure that a stable charger voltage
higher than 8-V is present at Pack+.
2. Press the Pack Voltage button to calibrate.
Figure 10. Calibration Screen
10 Pro (Advanced) Screen
10.1 SMB Communication
The set of read/write operations over SMBus are not specific to any gas gauge. These are provided as
general-purpose communication tools (Figure 11).
10.2 Hexadecimal/Decimal Converter
These two boxes convert between Hexadecimal value and Decimal value as soon as values are typed into
the boxes. Invalid values may cause erroneous results.