BiWaze® Clear System - User Manual Page | 42
Move the control unit away from any
sources of heat or hot ambient
temperatures. Switch off the device and
restart after 15 minutes of cool down.
If problem persists, call your health care
provider for servicing the device to replace
the filters and functional check.
Control unit does not power up Battery may be too
low.
Connect the control unit to the power cable
and check if the device powers on.
If problem persists, call your health care
provider for service.
Technical errors with an error
code number on a Red strip on
the LCD Screen and control unit
shuts down after few seconds
Technical error
related to
temperatures or other
high priority fault
Try rebooting the control unit and if
problem persists, call your health care
provider for service.
Technical errors with an error
code number on a Red strip on
the LCD Screen and control unit
does not shut down. User cannot
start the therapy.
Technical error
related to subsystem
malfunction
Try rebooting the control unit and if
problem persists, call your health care
provider for service.
Information with a self-
explanatory message on the LCD
screen in an Orange strip.
Informational
messages
User can acknowledge and continue with
therapy. Take action based on
informational message if needed.
Control unit not performing as
intended. Making abnormal
sounds or therapy performance.
Control unit
performance
malfunction.
Ensure that you move away from any high
electromagnetic or RF radiation sources
like MR machines, power transformers
etc.
If problem persists, do not use the control
unit and call your healthcare provider for
the service.
The following types of error messages along with their error codes may appear on-screen in case of a
malfunction.
Sr. No Notification Error Codes
1 Inspiratory Blower Error
1
2 Expiratory Blower Error
2
3 Insp. Pressure Sensor Error
4
4 Exp. Pressure Sensor Error
8
5 Insp. Flow Sensor Error
10
6 Exp. Flow Sensor Error
20
7 Barometric Pressure Sensor Error
40
8 Excess Pressure
80
9 High Delivered Air Temperature
100
10 High Battery Temperature
200
11 Positive Stepper Motor Error
400
12 Negative Stepper Motor Error
800
13 High MCB Temperature
1000
14 MCB Temperature Sensor Fail
2000
15 Delivered Air Temperature Sensor Fail
4000
16 Battery Temperature Sensor Fail
8000
17 Stepper Communication Error
10000
18 Pressure Sensor Mismatch
20000
19 Blower Calibration Error
40000
20 Flow Sensor Calibration Error
80000