User's Manual

BiWaze® Clear System - User Manual Page | 38
Transmit Power
1Mbps: 17.4 dBm (Typ)
54 Mbps: 13.8 dBm (Typ)
MCS7 (20MHz): 12.6 dBm (Typ)
MCS7 (40MHz): 11.3 dBm (Typ)
Receive Sensitivity
1Mbps DSSS: -96.3 dBm (Typ)
54 Mbps OFDM: -74.9 dBm (Typ)
MCS7 (20MHz): -72.4 dBm (Typ)
MCS7 (40MHz): -67.0 dBm (Typ)
Security Authentication/Encryption
Wi-Fi-protected access (WPA and WPA2.0) and IEEE Std
802.11i (includes hardware-accelerated Advanced Encryption
Standard [ AES ])
11.9 Control Accuracy
Parameter Accuracy
Pressure ± 5 cmH2O
Time ± 0.5 seconds
Frequency ± 1Hz
Device performance and accuracy is specified at Temperature: 20-30° C; Humidity: 50% relative; Altitude:
nominally 380 meters for typical patients.
11.10 Sound
The sound pressure of the device set at 30cmH2O in the PEP phase is less than 65 dBA at 1 meter.
11.11 Disposal
Dispose of this device and accessories in accordance with local regulations. This device and accessories
should be disposed of separately, not as unsorted municipal waste. To dispose of your device and accessories,
you should use the appropriate collection, reuse, and recycling systems available in your region. The use of
these collection, reuse, and recycling systems is designed to reduce pressure on natural resources and prevent
hazardous substances from damaging the environment
11.12 Essential Performance
The Essential Performance of the BiWaze Clear is defined as follows:
Therapy air pressure not to exceed 85 cmH2O.
12. EMC Information
WARNING - Portable RF communications equipment (including peripherals such as antenna cables
and external antennas) should be used no closer than 30 cm (12 inches) to any part of the BiWaze
Clear System, including cables specified by the manufacturer. Otherwise, degradation of the
performance of this equipment could result.”
WARNING - The BiWaze Clear System should not be used adjacent to or stacked with other equipment.
If adjacent or stacked use is necessary, the BiWaze Clear System should be observed to verify normal
operation. If operation is not normal, the BiWaze Clear System or the other equipment should be
moved.
WARNING - Use of accessories, transducers and cables other than those specified or provided by the
manufacturer of this equipment could result in increased electromagnetic emissions or decreased
electromagnetic immunity of this equipment and result in improper operation.
12.1 Electromagnetic Emissions
BiWaze Clear is intended for use in the electromagnetic environment specified below. The user of this device
should make sure it is used in such an environment.
Emissions Test Compliance Electromagnetic Environment - Guidance
RF emissions CISPR 11 Group 1 The device uses RF energy only for its internal