BiWaze® Clear System - User Manual Page | 34
8.4 Cleaning the Aerogen® Solo
The Aerogen® Solo Nebulizer is single patient use only and is not intended to be cleaned, disinfected or
sterilized.
8.5 Storage and transportation
While not in use cover the patient port with the cap provided at the port. Power off the control unit and remove
the power cable. Store in a dust free location outside the reach of children.
When in the home environment, use the carrying bag to transport or store the control unit. Do not place other
baggage on top of the device.
8.6 Traveling with BiWaze Clear
It may be helpful to bring the user manual along to help security personnel understand the device.
If you are traveling to a country with a line voltage different than the one you are currently using, a different
power cord or an international plug adaptor may be required to make your power cord compatible with the
power outlets of the country to which you are traveling.
8.7 Preventive Maintenance
This device does not require preventive maintenance.
9. Accessories
9.1. Carrying Bag
A carrying bag is available for BiWaze Clear when used in the home environment. When traveling, the carrying
bag is for carry-on luggage only. The carrying bag will not protect the system if it is put through checked
baggage.
9.2 Mobile Cart
A mobile cart is available for BiWaze Clear in the hospital environment. The mobile cart has a mounting plate,
a basket, and locking caster wheels. BiWaze Clear control unit is secured onto the mounting plate. The mobile
cart is an optional accessory. The cart provides mobility in acute and outpatient facilities.
10. Informational Messages
This chapter describes the informational messages that may appear on-screen and troubleshoots some of the
problems you may experience with your device and possible solutions to those problems.
The following type of informational messages may appear on-screen.
Message Description
Therapy complete Information Provides summary of the current therapy completion.
Error State Information In case of any technical errors, the Error Message is displayed.
11. Technical Specifications
11.1 Theory of operations
BiWaze Clear is designed around a pneumatic assembly that controls positive as well as negative pressure
and flow delivery to the patient. The main processor monitors sensors for pressure, flow and so on, and
controls the blowers to meet treatment settings and make breathing comfortable for the user. A number of
internal sensor readings are monitored to ensure that the BiWaze Clear functions correctly. Some of them are
checked at power up, some at therapy start, and some are monitored continuously.
Main Control Board (MCB): This board has multiple processors including main processor for control of
pressure and flow. This board controls the positive and negative flow control valves, blowers as well as
monitors various temperatures and battery capacity. It also communicates with UI board and Sensor board.