BiWaze® Clear System - User Manual Page | 18
10. Once the therapy is running, the
therapy can be paused at any time
by pressing the Pause button or
the Therapy Mode Light.
Note: The therapy can be paused for 5
minutes. A countdown timer will
appear on the screen. You can choose
to resume therapy or end therapy. If
the timer reaches 5 minutes the
therapy will automatically end.
11. When you have completed the
Manual therapy, press the Pause
button and then select END
NOW.
12. A ribbon will be displayed across
the screen letting you know the
Therapy is Complete.
a. Press the to view the
therapy summary.
b. Press the on the left to
close the ribbon.
5.3.2 Edit Manual Therapy
Manual therapy can be edited if the software is unlocked. Typically, the software will be locked when the
BiWaze Clear is in the home environment. See Section 7 Locking and Unlocking Feature for more information
about locking and unlocking the software.
The editable parameters include the following:
1.
To edit PEP, press the phase circle
2.
Press the up arrow to increase the
pressure and the down arrow to
decrease the pressure
3.
Press the checkmark to save the
parameter and return to the therapy
screen
4.
To edit OSC, press the phase circle
5.
Press the up and down arrows to edit
the pressure
6.
Press next to the frequency you want
to deliver: Low, Medium or High
7.
Press the checkmark to save the
parameters and return to the therapy
screen
Parameter Description
PEP Pressure Set the peak pressure from 0 to 30 cmH
2
O in increments of 1.
OSC Pressure Set the peak pressure from 5 to 70 cmH
2
O in increments of 1.
OSC Frequency
Set the frequency of oscillations to be Low (3 hertz), medium (4 hertz) or high (5
hertz) for this phase.