User's Manual
BiWaze® Clear System - User Manual Page | 14
8. When the therapy is complete, a
ribbon will be displayed across
the screen.
8.1. Press the to view the
therapy summary.
8.2. Press the on the left to
close the ribbon.
5.2.1 Edit Auto Therapy
Auto therapy can be edited if the system is unlocked or if Allow User Adjustment is enabled under Lock
Options. Typically, healthcare providers will lock the software when BiWaze Clear is used in the home
environment. See Section 7 Locking and Unlocking Feature for more information about locking and unlocking
the software.
The editable parameters include the following:
Parameter Description
PEP Pressure Set the peak pressure from 0 to 30 cmH
2
O in increments of 1.
PEP Time Set the phase time from 0:05 to 30:00 minutes in increments of 5 seconds.
PEP Neb Set the nebulizer to be on or off during this phase of therapy.
OSC Pressure Set the peak pressure from 5 to 70 cmH
2
O in increments of 1.
OSC Time Set the phase time from 0:05 to 30:00 minutes in increments of 5 seconds.
OSC Frequency
Set the frequency of oscillations to be Low (3 hertz), medium (4 hertz) or high (5
hertz) for this phase.
OSC Neb Set the nebulizer to be on or off during this phase of therapy.
NEB
Set the phase time from 0:05 to 30:00 minutes in increments of 5 seconds. This
phase is for running the nebulizer only, no adjustment in pressure or oscillations.