For intermittent doses less than or equal to 6 mL:
1. Open the medication port cap on the Aerogen®
Solo
2. Use a pre-filled ampoule or syringe to add
medication into the filler port of the Aerogen®
Solo
3. Close the port cap
Note: Medication can be added to the Aerogen®
Solo during therapy.
3.4. Using In-line with a Ventilator
To prevent injury or equipment damage, read these warnings before assembly
WARNING - Always use a new Dual Lumen Breathing Circuit when using the system on a new patient
to avoid cross-contamination.
WARNING - Only trained person on BiWaze Clear and the ventilator should provide the therapy to
ventilated patients.
1. Follow the steps in Section 3.1
Connect the components
2. Connect a spring valved adapter on
the inspiratory limb of the ventilator’s
dual limb breathing circuit
3. Connect the patient side of the
handset on to the spring valved
adapter
4. Power On and Off the Control Unit
The control unit can operate on external AC power or internal battery when charged.
4.1 Using AC Power
4.2 Internal Battery
BiWaze Clear comes with an internal Lithium-ion battery pack. The battery can power the device with active
therapies for up to 2 hours* on full charge.
The internal battery can charge simultaneously while the device is operating on the AC power and switches to
battery power source when AC power is disconnected.
*Subjected to default settings, the actual run time can vary depending on age of the battery, settings and actual
active therapy time.
Note: The battery shall be fully charged before using the device for the first time or when device is unused for
an extended period of time.
Note: Do not dispose the battery in landfill.
WARNING - Only authorized service personnel should replace the battery in the control unit.
An AC power cord is included with the control unit.
1. Plug the socket end of the power cord (A) into the AC inlet on the
back of the control unit.
2. Plug the pronged end (B) into an electrical outlet not connected to
a wall switch.
Note: Ensure all connections are secure.