User's Manual

BiWaze® Clear System - User Manual Page | 8
Trach Adapter
i. Connect the handset (C) to the flexible trach
adapter (G)
ii. Connect the other end of the flexible trach
adapter to the patient’s tracheostomy tube
3.2. Connect to an Oxygen Source
To prevent injury or equipment damage, read these warnings before assembly
WARNING - Always use a new Dual Lumen Breathing Circuit when using the system on a new patient
to avoid cross-contamination.
WARNING - Do not connect the system to an unregulated or high-pressure oxygen source.
CAUTION - When using oxygen with this system, power on the system before you turn on the oxygen
source. Before you power down the system, turn off the oxygen source. This helps prevent oxygen
from accumulating in the system.
1. Follow the steps in Section 3.1
Connect the components
2. Connect the oxygen tubing from the
oxygen source to the oxygen port
present in the B/V Filter
3.3. Aerogen® Solo Nebulizer Assembly and Add Medication
To prevent injury or equipment damage, read these warnings before assembly
WARNING - Only use Aerogen® Solo nebulizer with BiWaze Clear.
WARNING - Always use a new Dual Lumen Breathing Circuit when using the system on a new patient
to avoid cross-contamination.
WARNING - Aerogen® Solo is a single patient use device not to be used on more than one patient
to prevent cross-infection.
WARNING - The components and accessories of the Aerogen® Solo, as packaged, are not sterile.
WARNING - Do not autoclave any component or accessory of the Aerogen® Solo System.
WARNING - Do not use or store outside of specified environmental conditions.
Note: The components and accessories of the Aeroge Solo are not made with natural rubber latex.