BiWaze® Clear System - User Manual Page | 7
3. Assemble The Dual Lumen Breathing Circuit
• WARNING - Do not operate the system without B/V filter
• WARNING - Do not apply excessive force to assemble or disassemble the Dual Lumen Breathing
Circuit to avoid damage to the components.
• WARNING - Always use a new Dual Lumen Breathing Circuit when using the system on a new
patient to avoid cross-contamination.
• WARNING - Do not sterilize the circuit for reuse.
Note: Each Dual Lumen Breathing Circuit is for use by a single patient and intended for 30 days of treatment
or a maximum of 90 treatment sessions.
3.1. Connect the components
1.
Attach the larger end of the coaxial tube (B) to
the end of the B/V filter (A) that has the
company logo.
2.
Attach the end of the B/V filter (A) that has the
oxygen port to the patient port of the control
unit.
3. Connect the smaller end of the coaxial tube (B)
to the end of the handset (C) closest to the
nebulizer port.
4. Attach the applicable patient interface:
• Face Mask
i.Insert and gently twist the face mask (E) to the
handset (C) patient interface end
• Mouthpiece
Insert and gently twist the mouthpiece(F) with
the handset (C).