Datasheet

www.ti.com
TEST AFE5809
Figure 7. TSW1400 GUI Configuration
From the AFE5809 GUI, go to the ADC page, and then select Ramp.
Press the Capture button on the TSW1400 GUI. This displays a ramping waveform on the TSW1400
GUI display area as shown in Figure 8.
Verify that Maximum value is 16383 for 14-bit mode.
Repeat for Channel 2 and Channel 8.
If each channel has the output as shown in Figure 8, proceed with the next step; otherwise, contact the
TI FAE to troubleshoot the problem.
On the AFE5809 GUI, change Test Pattern from Ramp to None for the next step.
9
SLOU338AOctober 2012Revised November 2012 AFE5809EVM Evaluation Module
Submit Documentation Feedback
Copyright © 2012, Texas Instruments Incorporated