Hardware Setup, CW Mode
9.3 External Clock for CW Mode
The CW mode clocks can be supplied from J9 (X16/X8/X4/X1) and J10 (X1) using external signal
generators. The manufacture default setup uses ON BOARD CLOCK, with which C154, C155, C156, and
C157 are installed. To switch to an external clock, these four capacitors must be uninstalled and
capacitors C49,C50,C52, and C53 must be installed. The following table shows the configuration for these
eight capacitors.
Table 1. Configuration for Clock Capacitors
Capacitor Onboard Clock External Clock Comments
C27 Uninstalled Installed Top layer of the EVM
C28 Uninstalled Installed Bottom layer of the EVM
C100 Uninstalled Installed Bottom layer of the EVM
C102 Uninstalled Installed Top layer of the EVM
C29 Installed Uninstalled Top layer of the EVM
C30 Installed Uninstalled Top layer of the EVM
C25 Installed Uninstalled Top layer of the EVM
C26 Installed Uninstalled Top layer of the EVM
Figure 22. Relevant Capacitors for CW Mode, Top and Bottom Side
9.4 External ADC Sampling Clock
To use the external clock generator to test the AFE5809, perform the following steps.
1. Reconfigure JP9 as shown in Figure 23. The rest of the jumpers remain the same.
2. Connect the external generator as shown in Figure 23.
3. Set the generator output to 65 MHz, 13 dBm.
23