9.2 Step 2: Apply Input and Observe Outputs
• Apply an analog signal (2.51 M, –10 dBm) to any analog input SMA.
• The CW outputs (J12, J13) display the frequency I and Q signals at 10 kHz as shown in Figure 21.
The GUI Gain Control Feedback Resistor can be used to vary the amplitude of the outputs.
Figure 21. CW Outputs