Datasheet
TEST AFE5809
www.ti.com
Figure 8. Ramp Waveform Output
8.2 Step 2: Single Tone FFT
• Select Frequency from the display control on the TSW1400 GUI to display the FFT.
• Connect Channel 1 of the AFE5809EVM to a signal generator through a BP filter. If a BP filter is not
present, the result is not good.
• Set the amplitude of the signal generator to -25 dBm.
• Set the frequency of the signal generator to the value in the "ADC Input Target Frequency" field of the
TSW GUI.
• Change the window option to Hanning if using the onboard clock. This is because the input signal and
the onboard CMOS clock are noncoherent.
• Press the Capture button to get the test result.
• Repeat for Channel 2 through Channel 8.
10
AFE5809EVM Evaluation Module SLOU338A–October 2012–Revised November 2012
Submit Documentation Feedback
Copyright © 2012, Texas Instruments Incorporated