8.2 Step 2: Single Tone FFT
• Select the Single Tone FFT page at the TSW1250 GUI.
• Connect Channel 1 of the AFE5803EVM to a signal generator through an LP filter. If an LP filter is not
present, the result will not be good.
• Set the amplitude of the signal generator to -25dBm.
• Set the frequency of the signal generator to 2.00439453M to match the GUI.
• Change the window option to Hanning. This is because the input signal and the onboard CMOS clock
are noncoherent.
• Press the Capture button to get the test result.
• Repeat for Channel 2 through Channel 8.