User manual
www.ti.com
AFE44x0SPO2EVM FAQs
10 AFE44x0SPO2EVM FAQs
10.1 EVM Communicating With the PC Application
CAUTION
AFE44x0SPO2EVM GUI v.2.0 works with FW revision 1.3. Follow the steps
outlined in Section 6 to upgrade the firmware to revision 1.3.
A quick and simple check to verify serial register write operation is to put the AFE44x0 in power-down
mode. Follow the sequence to check if the GUI is communicating with the EVM.
• In Device Configuration→Global Settings tab, select Powerdown_AFE
• This powers down the AFE and the VCM output voltage of the AFE drops to 0 V
• VCM is measured at the VCM_AFE serial jumper resistor R28 on the board
10.2 ADC_RDY Signal
After executing the GUI, observe the ADC_RDY waveform at series jumper resistor R26.This should be at
the same frequency as the PRF. Figure 45 shows the ADC_RDY waveform at 500-Hz PRF.
Figure 45. ADC_RDY Waveform at 500-Hz PRF
37
SLAU480C–January 2013–Revised May 2014 AFE4400 and AFE4490 Development Guide
Submit Documentation Feedback
Copyright © 2013–2014, Texas Instruments Incorporated