User manual

www.ti.com
Running the Software
4. No. of samples = 30
Data rate (sps) = 500
Show data for the last 1 secs
Then display Error message “Insufficient # of samples for FFT calculation” since # of samples
for FFT which is power of 2 = 16 min ( (500 x 1) , 30 )
5. No. of samples = 32
Data rate (sps) = 500
Show data for the last 2 secs
# of samples for FFT which is power of 2 = 32 min ( (500 x 2) , 32 )
Since (# of samples for FFT calc. == No. of samples) and If (Filter Type = None) then # of
samples for FFT calc. which is power of 2 = 32
Since (# of samples for FFT calc. == No. of samples) and If (Filter Type = Notch) then an error
msg Insufficient # of samples for FFT calculation” will be displayed since # of samples for FFT
calc. which is power of 2 = (32 / 2) < 32 samples
Figure 31. ADC Capture and Analysis Tab
Figure 32. Scope Analysis: Test Results
25
SLAU480CJanuary 2013Revised May 2014 AFE4400 and AFE4490 Development Guide
Submit Documentation Feedback