User manual
www.ti.com
Running the Software
Figure 27. AFE4400: Device Configuration: Rx Stage
4.1.1.4 Timing Controls Subtab
The Timing Controls subtab under the Device Configuration tab, shown in Figure 28 for AFE4490 and
Figure 29 for AFE4400, consists of the following settings:
1. Enter the Pulse Repetition Frequency (PRF) and Duty Cycle % and click the SET button to
automatically set the following:
(a) LED1 (IR) and LED2 (Red) ON and OFF time,
(b) Rx sample start and end time for 4 channels (LED1, LED1 Ambient, LED2, LED2 Ambient)
(c) Rx convert start and end time for 4 channels (LED1, LED1 Ambient, LED2, LED2 Ambient)
2. Save the timing settings based on PRF and duty cycle to a configuration file
3. Load the timing settings based on PRF and duty cycle from a configuration file
4. Timer Enable selector
5. Timer Counter RESET button
6. Set Number of Averages (available for AFE4490 device only)
21
SLAU480C–January 2013–Revised May 2014 AFE4400 and AFE4490 Development Guide
Submit Documentation Feedback
Copyright © 2013–2014, Texas Instruments Incorporated