Datasheet
AFE4490
www.ti.com
SBAS602F –DECEMBER 2012–REVISED OCTOBER 2013
RECOMMENDED OPERATING CONDITIONS
Over operating free-air temperature range, unless otherwise noted.
PARAMETER VALUE UNIT
SUPPLIES
RX_ANA_SUP AFE analog supply 2.0 to 3.6 V
RX_DIG_SUP AFE digital supply 2.0 to 3.6 V
TX_CTRL_SUP Transmit controller supply 3.0 to 5.25 V
H-bridge or common [3.0 or (V
HR
+ V
LED
+ V
CABLE
)
(1)(2)(3)
,
LED_DRV_SUP Transmit LED driver supply V
anode configuration whichever is greater] to 5.25
Difference between LED_DRV_SUP and TX_CTRL_SUP –0.3 to +0.3 V
TEMPERATURE
Specified temperature range –40 to +85 °C
Storage temperature range –60 to +150 °C
(1) V
HR
refers to the required voltage headroom necessary to drive the LEDs. See Table 7 for the appropriate V
HR
value.
(2) V
LED
refers to the maximum voltage drop across the external LED (at maximum LED current) connected between the TXP and TXN pins
(in H-bridge mode) and from the TXP and TXN pins to LED_DRV_SUP (in the common anode configuration).
(3) V
CABLE
refers to voltage drop across any cable, connector, or any other component in series with the LED.
Copyright © 2012–2013, Texas Instruments Incorporated Submit Documentation Feedback 3
Product Folder Links: AFE4490