8 Quick Start Guide
NOTE: For all the measurements shown below, ProSim Fluke SPOT Light SpO2 Functional tester
was used.
Install the AFE44x0SPO2EVM GUI PC Software (Check the TI website at www.ti.com for the latest
software) by running the setup file and following the instructions on the screen
Connect the USB cable to the computer and plug in the mini USB interface cable to the J4 mini USB
connector on the EVM
Connect the DB9 connector of the finger sensor pulse oximeter cable to the J2 connector of the EVM.
Turn on the Fluke simulator by pressing the ON switch for at least three seconds and then connect the
Fluke simulator to the finger sensor. Leave the default setting of 97% for SpO2, 80 bpm for HR, and
2% for PA. The Fluke SPOT Light Pulse SpO2 Functional Tester setup with the finger sensor pulse
oximeter cable is shown in Figure 30
Open the AFE44x0SPO2EVM GUI software by clicking on StartAll ProgramsTexas
InstrumentsAFE44x0SPO2EVM GUI.
Click the ADC Capture & Analysis tab and set the following:
Capture Mode to Finite
No. of Samples to 2048
Volts/Codes to Volts
Plot Mode to Four Plot Mode
Select the first waveform to LED1 (IR)
Select the second waveform to LED1 (IR) AMBIENT
Select the third waveform to LED2 (Red)
Select the fourth waveform to LED2 (Red) AMBIENT
Click Capture
Figure 38 shows a sample data capture
33
SLAU480AJanuary 2013Revised February 2013 AFE4400 and AFE4490 Development Guide
Submit Documentation Feedback