Datasheet

www.ti.com
Connector Interface
Figure 35. PC Application Firmware Upgrade 5
7 Connector Interface
The following connectors are used for external interface to the AFE44x0 Pulse Oximeter board.
DB9
USB mini connector
7.1 DB9 Pulse Oximeter Connector
The DB9 pulse oximeter connector pin-outs are shown in Figure 36. The description of the pin-outs is
provided in Table 6
Figure 36. DB9 Pulse Oximeter Connector Pin Outs
Table 6. DB9 based Pulse Oximeter Connector pinouts
Pin Number Pin Name Pin Description
2 TX_LED_P Anode of the IR LED, cathode of the red LED
3 TX_LED_N Cathode of the IR LED, anode of the red LED
5 DET_N Phototransistor anode
7 GND Cable shield
9 DET_P Phototransistor cathode
31
SLAU480AJanuary 2013Revised February 2013 AFE4400 and AFE4490 Development Guide
Submit Documentation Feedback
Copyright © 2013, Texas Instruments Incorporated