Datasheet
Advanced bqMTester Instructions
www.ti.com
4.6.8 Voltage Frame
This frame will be grayed out if not selected in the Type of calibration to perform frame. This frame
contains two values:
1. Reference/FSV: The tester calibrates the voltage gain by manipulating the Full Scale Voltage
Reference. Do not change the values in this field. The Voltage frame looks the same for most of the
parts, except for the bq306x. For bq306x device-based parts, extra K-factor fields appear. The frame
for K-factor–based parts are shown in Figure 7.
Figure 7. Reference/FSV and K-Factor Fields
2. % Error: The %Error field is used as a rough test to check the Voltage Measurement Circuitry. After
the bqMTester calibrates the voltage gain, then it compares the new calibration value to what is in the
Reference/FSV field.
4.6.8.1 Limit to Force the Voltage Calibration
If the percent difference between the two values is more than 25%, then it fails the calibration because it
assumes something must be grossly wrong to get a value more than 25% from the nominal. As shown in
Figure 7, bq306x device-based parts show limits to voltage calibration in ±mV. In this case, if the voltage
read by the device is within this limit, no calibration is done. Precious time is saved (~4s) during calibration
if voltage calibration can be avoided. In the case where voltage calibration is always needed, a 0 can be
entered in this field.
4.6.9 Temperature Frame
The Temperature frame will be grayed out if not selected in the Type of calibration to perform frame.
This frame contains one value. Enter the maximum absolute value of offset that the bqMTester software
will be allowed to put into any of the data flash temperature offset registers for the tested module. This is
not an accuracy verification. This is a gross Error detection. The default value of this field is 40, meaning
that the calibrated offset put in the data flash cannot exceed positive or negative 4°C. For internal
Temperature Sensor calibration, it is recommended to increase this value because the internal
temperature sensor offset accuracy commonly exceeds 4°C.
4.6.10 Starting Serial Number
Enter the value for the serial number of the first smart battery module to be tested. This number will be
incremented by one as each new module is tested. If Skip On Error is checked, the number will not
increment in the case of a module that fails the test. The default for this box is 1. This value must be
specified as a positive integer value.
4.6.11 Date
Enter the value for the desired date to be programmed into the smart battery module. If Use Current Date
is checked, the system date from the PC running the bqMTester software will be used.
4.6.12 Log File Name
Enter the complete path and file name to be used for the log file. This file contains all relevant test data for
each smart battery module tested. If Clear Log button is clicked, the log file contents will be deleted.
4.6.13 Pack Lot Code or Manufacturer Info
Enter the value for the Lot Code of the group of smart battery modules currently being tested. This
number will not change until it is changed manually and will be programmed into each smart battery
module tested. This value must be specified as a positive integer value.
20
Advanced bqMTester SLUU397A–February 2011– Revised May 2011
Submit Documentation Feedback
Copyright © 2011, Texas Instruments Incorporated