Datasheet

Basic Test Procedure
www.ti.com
Figure 12. ADS58H4x 11-bit SNRBoost Mode, Single Tone Test Result: Fs = 245.76 MSPS, Fin = 170 MHz
Single tone FFT test (ADS58H4x 14-bit High Resolution Burst Mode)
1. Select the ADS58H40-14b device.
2. After selecting Single Tone FFT Test under Test Selection, enter the sample rate, input frequency, and
the number of samples.
3. Select channel A, B, C, or D depending on which channel is connected to the signal generator.
4. On the High Speed Data Converter Pro menu, select Data Capture Options Trigger Option. Check
both Trigger Mode Enable and Software Trigger Enable.
14
ADS58H4x (ADS58H40/ADS58H43) EVM SLAU455AAugust 2012Revised October 2012
Submit Documentation Feedback
Copyright © 2012, Texas Instruments Incorporated