User's Manual
Table Of Contents

www.ti.com
* dBFS + 20 log
captured max code * captured min code
2
N
, where N is the number of bits.
EVM Operational Procedure
Table 1-1. Three Pin Jumper List Table
JUMPER FUNCTION LOCATION: PINS 1–2 LOCATION: PINS 2–3 DEFAULT
SJP3 Provides AIN+ source to ADC Source provided from T2 Source provided from Diff Amp 1–2
device
SJP4 Provides AIN– source to ADC Source provided from Diff Amp Source provided from T2 2-3
device
2. Connect supplies to the EVM as follows:
+5 V (4.75 V–5.25 V) ADC analog supply to J3 and return to J2.
+3.3 V (3 V–3.6 V) digital buffer supply to J4 and J1 and return to J6
3. Switch power supplies on.
4. Use a function generator with 50- Ω output to input a 105-MHz, 0-V offset, 1-Vrms sine-wave signal into
J5. The frequency of the clock must be within the specification for the device speed grade.
5. Supply an input signal by using a frequency generator with a 50- Ω output to provide a 15.5 MHz, 0-V
offset, –1-dBFS amplitude sine-wave signal into J11. A full-scale input tone into the ADC device is a
differential 2.2 Vpp and dBFS can be calculated by using the following formula:
6. The digital pattern on the output connector J9 should now represent a 2's compliment sine wave and
can be monitored using a logic analyzer.
6 Overview SLWU020B – February 2005 – Revised February 2006
Submit Documentation Feedback