Datasheet
Test Setup
www.ti.com
4 Test Setup
To evaluate the ADS5295 device, a TSW1400EVM is required. Figure 3 shows the exact setup of these
two boards and external connectors.
• Connect the P10 connector of the ADS5295EVM to the J3 connector of the TSW1400EVM.
• Connect a +5-V supply at the P1 connector or across the TP_VP (+5 V) and TP2 (GND).
• Connect a +5-V wall adapter to the TSW1400EVM at the J12 connector.
• Connect a USB cable from the PC to USB1 on the ADS5295EVM.
• Connect a USB cable from the PC to USB port J5 of the TSW1400EVM.
• An external sampling clock is not required as the onboard 80-MHz CMOS clock oscillator is used in the
default configuration.
Figure 3. Hardware Setup of TSW1400EVM and ADS5295EVM
12
ADS5295, 8-Channel, Analog-to-Digital Converter Evaluation Module SLAU442–November 2012
Submit Documentation Feedback
Copyright © 2012, Texas Instruments Incorporated