Datasheet

www.ti.com
Using the Software: ADS1299 Control Registers and GUI
Figure 19. Register Bit for SRB1 Routing
5.3.1 Internal Test Signals Input
Configuration Register 2 controls the signal amplitude and frequency of an internally-generated square
wave test signals. The primary purpose of this test signal is to verify the functionality of the front-end MUX,
the PGA, and the ADC. The test signals may be viewed on the AnalysisScope tab, as Figure 20 shows.
Detailed instructions for using the AnalysisScope tab is provided in Section 6.1.1.
Figure 20. Internal Test Signals
5.3.2 Temperature Sensor and the Scope Tab
The internal temperature sensor on the ADS1299 is shown in . When the internal MUX is routed to the
temperature sensor input, the output voltage of the ADC may be converted to a temperature value, using
Equation 1.
(1)
21
SLAU443May 2012 EEG Front-End Performance Demonstration Kit
Submit Documentation Feedback
Copyright © 2012, Texas Instruments Incorporated