Datasheet

ADS1294, ADS1294R
ADS1296, ADS1296R
ADS1298, ADS1298R
SBAS459I JANUARY 2010REVISED JANUARY 2012
www.ti.com
RESPIRATION
The ADS1294R/6R/8R provide three modes for respiration impedance measurement: external respiration,
internal respiration using on-chip modulation signals, and internal respiration using user-generated modulation
signals, as shown in Table 14.
Table 14. Respiration Control
RESP.RESP_CTRL[1] RESP.RESP_CTRL[0] MODE AVAILABLE DESCRIPTION
ADS1294, ADS1296, ADS1298, Respiration disabled
0 0
ADS1294R, ADS1296R, ADS1298R
Generates modulation and demodulation signals for external
ADS1294, ADS1296, ADS1298,
0 1 respiration circuitry. RESP CLK signals on GPIO2, GPIO3, and
ADS1294R, ADS1296R, ADS1298R
GPIO4.
Respiration measurement using internally-generated
1 0 ADS1294R, ADS1296R, ADS1298R
RESP_MOD signals.
ADS1294R, ADS1296R, Respiration measurement using user-generated modulation and
1 1
ADS1298R
(1)
blocking signal.
For more information on respiration impedance measurement, refer to application note Respiration Rate
Measurement Using Impedance Pneumography (SBAA181).
(1) RESP_CTRL[1:0] = 11 is not recommend if CLKSEL = 1 (internal master clock).
70 Submit Documentation Feedback Copyright © 20102012, Texas Instruments Incorporated
Product Folder Link(s): ADS1294 ADS1294R ADS1296 ADS1296R ADS1298 ADS1298R