OVERVIEW
NOTE
The ADS1294R/6R/8R channel performance differs from the ADS1294/6/8 in regards to
respiration circuitry found on channel one. Unless otherwise noted, ADS129x refers to all
specifications and functional descriptions of the ADS1294, ADS1296, ADS1298,
ADS1294R, ADS1296R, and ADS1298R.
The ADS129x are low-power, multichannel, simultaneously-sampling, 24-bit delta-sigma (ΔΣ) analog-to-digital
converters (ADCs) with integrated programmable gain amplifiers (PGAs). These devices integrate various
ECG-specific functions that make them well-suited for scalable electrocardiogram (ECG),
electroencephalography (EEG), and electromyography (EMG) applications. The devices can also be used in
high-performance, multichannel data acquisition systems by powering down the ECG-specific circuitry.
The ADS129x have a highly programmable multiplexer that allows for temperature, supply, input short, and RLD
measurements. Additionally, the multiplexer allows any of the input electrodes to be programmed as the patient
reference drive. The PGA gain can be chosen from one of seven settings (1, 2, 3, 4, 6, 8, and 12). The ADCs in
the device offer data rates from 250SPS to 32kSPS. Communication to the device is accomplished using an
SPI-compatible interface. The device provides four GPIO pins for general use. Multiple devices can be
synchronized using the START pin.
The internal reference can be programmed to either 2.4V or 4V. The internal oscillator generates a 2.048MHz
clock. The versatile right leg drive (RLD) block allows the user to choose the average of any combination of
electrodes to generate the patient drive signal. Lead-off detection can be accomplished either by using a
pull-up/pull-down resistor or a current source/sink. An internal ac lead-off detection feature is also available. The
device supports both hardware PACE detection and software PACE detection. The Wilson Central Terminal
(WCT) block can be used to generate the WCT point of the standard 12-lead ECG.
Additionally, the ADS1294R, ADS1296R, and ADS1298R provide options for an internal respiration modulator
and a demodulator circuit in the signal path of channel 1.