Datasheet
ADS1293
SNAS602B –FEBRUARY 2013–REVISED MARCH 2013
www.ti.com
Table 38. ERROR_STATUS: Other Error Status
Addr BIT7 BIT6 BIT5 BIT4 BIT3 BIT2 BIT1 BIT0
0x19 SYNC EDGEERR CH3ERR CH2ERR CH1ERR LEADOFF BATLOW RLDRAIL CMOR
[7] SYNCEDGEERR Digital synchronization error
1: Indicates a digital synchronization error occurred
[6] CH3ERR Channel 3 out-of-range error
1: Indicates an out-of-range error detected on channel 3
[5] CH2ERR Channel 2 out-of-range error
1: Indicates an out-of-range error detected on channel 2
[4 ] CH1ERR Channel 1 out-of-range error
1: Indicates an out-of-range error detected on channel 1
[3] LEADOFF Lead off detected
1: Indicates a lead off was detected on at least one input pin
[2] BATLOW Low battery
1: Indicates the battery voltage has dropped below 2.7 V
[1] RLDRAIL Right leg drive near rail
1: Indicates the right leg drive amplifier output is approaching the supply rails
[0] CMOR Common-mode level out-of-range
1: Indicates the level detected by the common-mode detect block is outside of the input common-
mode range of the amplifiers in the analog front-end
Note: The clock to digital (internal or external) must be enabled in 0x12[2] for this error register to update.
Table 39. ERROR_RANGE1: Channel 1 AFE Out-of-Range Status
Addr BIT7 BIT6 BIT5 BIT4 BIT3 BIT2 BIT1 BIT0
0x1A SDM_ OR_CH1 SIGN_CH1 OUTN_ OUTN_ OUTP_ OUTP_ DIF_HIGH_
LOW_CH1 HIGH_CH1 LOW_CH1 HIGH_CH1 CH1
[7] RESERVED —
[6] SDM_OR_CH1 Channel 1 sigma-delta modulator over range
1: Indicates an over range detected for channel 1 SDM
[5] SIGN_CH1 Channel 1 instrumentation amplifier output sign
This bit specifies the sign of the output signal of the instrumentation amplifier for channel 1.
0: Positive output of INA larger than negative output
1: Positive output of INA smaller than negative output
[4 ] OUTN_LOW_CH1 Channel 1 instrumentation amplifier negative output near negative rail
1: Indicates the negative output of the INA is close to the negative rail for channel 1
[3] OUTN_HIGH_CH1 Channel 1 instrumentation amplifier negative output near positive rail
1: Indicates the negative output of the INA is close to the positive rail for channel 1
[2] OUTP_LOW_CH1 Channel 1 instrumentation amplifier positive output near negative rail
1: Indicates the positive output of the INA is close to the negative rail for channel 1
[1] OUTP_HIGH_CH1 Channel 1 instrumentation amplifier positive output near positive rail
1: Indicates the positive output of the INA is close to the positive rail for channel 1
[0] DIF_HIGH_CH1 Channel 1 instrumentation amplifier output out-of-range
1: Indicates the differential output voltage of the INA is out-of-range for channel 1
Note: The clock to digital (internal or external) must be enabled in 0x12[2] for this error register to update.
60 Copyright © 2013, Texas Instruments Incorporated
Product Folder Links: ADS1293