Datasheet

MUX1[3:0] = 0000
RLD1P = 1
RLD1N = 1
MUX1[3:0] = 0111
PGA1
RLD2P = 0
RLD2N = 0
PGA2
RLD_AMP
RLDOUT RLDINVRLDIN/RLDREF
Filter or
Feedthrough
ADS1292R
EMI
Filter
EMI
Filter
IN1P
IN1N
IN2P
IN2N
MUX
1 M
(1)
1.5 nF
(1)
2
(AVDD + AVSS)
RLDREF_INT = 1
RLDREF_INT = 0
ADS1291
ADS1292
ADS1292R
SBAS502B DECEMBER 2011REVISED SEPTEMBER 2012
www.ti.com
ECG-SPECIFIC FUNCTIONS
INPUT MULTIPLEXER (REROUTING THE RIGHT LEG DRIVE SIGNAL)
The input multiplexer has ECG-specific functions for the right leg drive signal. The RLD signal is available at the
RLDOUT pin once the appropriate channels are selected for RLD derivation, feedback elements are installed
external to the chip, and the loop is closed. This signal can be fed after filtering or fed directly into the RLDIN pin,
as shown in Figure 47. This RLDIN signal can be multiplexed into any one of the input electrodes by setting the
MUX bits of the appropriate channel set registers to '0110' for P-side or '0111' for N-side. Figure 47 shows the
RLD signal generated from channel 1 and routed to the N-side of channel 2. This feature can be used to
dynamically change the electrode that is used as the reference signal to drive the patient body. Note that the
corresponding channel cannot be used and can be powered down.
(1) Typical values for example only.
Figure 47. Example RLDOUT Signal Configured to be Routed to IN2N
50 Submit Documentation Feedback Copyright © 2011–2012, Texas Instruments Incorporated
Product Folder Links: ADS1291 ADS1292 ADS1292R