Datasheet
www.ti.com
Evaluating Performance with the ADCPro Software
The Power Supply control selects between single supply (+5V) mode and bipolar supply (±2.5V) mode.
The EVM draws its supply voltages from the MMB3 motherboard, and has onboard circuitry to select
between the +5V and bipolar supplies.
The Reference Select control selects between VREF0, VREF1, and the Internal Reference. When
VREF1 is selected, the VREF value is set to 2.048V, with switch S1 set to BUFF. S1 should not be set to
the AVDD position, in order to prevent the reference from exceeding the maximum specified input.
There are two checkboxes that, when enabled, allow the REF0P and REF0N pins to be selected as
GPIOs. These checkboxes are enabled only in modes that allow these pins to be used as GPIOs.
The Internal Reference control selects whether the reference is off, on, or follows the START signal (refer
to the ADS1248 data sheet for more details on these choices). When the internal reference is enabled, the
IDAC controls are enabled as well, because the IDACs cannot function unless the internal reference is
turned on. Note that the internal reference does not necessarily have to be used as the ADC reference in
order to use the IDACs, but it does need to be turned on.
The IDAC Magnitude control sets the amount of current each of the two IDACs provide.
The IDAC0 and IDAC 1 Routing controls allow you to specify where the IDAC0 or IDAC1 current go: to
either the IEXC1 or IEXC2 pins, or to the Analog inputs, or you may turn the IDAC off. If the AIN button is
selected, this option enables the corresponding IDAC controls on the I/O Config tab.
25
SBAU142B– April 2009–Revised May 2011 ADS1148EVM, ADS1248EVM, ADS1148EVM-PDK, and ADS1248EVM-PDK
Submit Documentation Feedback
Copyright © 2009–2011, Texas Instruments Incorporated