Datasheet
FIR
Filters for
ECG (CH2)
& Respiration (CH1)
Data Acquisition
From ADC
ADC data DRDY
Interrupt
MemoryUSB
PC Application
SPI Interrupt
2 ChannelData
Reader
IIR
Filter
(DC Removal)
QRS
Algorithm
(HR) CH2
Lead Off
Status
Respiration
Rate
Algorithm
(RR) CH1
ECG signal
Respiration signal
ADS1292R DATA FLOW DIAGRAM:
PROCESSED ECG and RESPIRATION
DRDY
Interrupt
Decode PC
Commands
Lead Off Status
Lead Off Status
HR
1 ECG Lead
1 RESPIRATION CH.
Lead Off Status
RR
USB
SPI
-1
-1
Y(z) 1 - z
H(z) = =
X(z) 1 - az
www.ti.com
Evaluation of ECG and Respiration Specific Functions
Figure 37. Live Data Stream Mode Data Flow for ADS1292R
6.2.1 Filters
IIR Filter - DC signal removal
1st Order IIR filter is used to remove the DC component from the acquired data. The following transfer
function is used for the filter:
(1)
To provide DC attenuation at 22dB, the value of alpha is chosen as 0.992. The IIR filter output is scaled
down to 16-bit and then provided to the FIR filter. The frequency response for the IIR filter is shown in
Figure 38.
33
SLAU384A–December 2011–Revised April 2012 ADS1x9xECG-FE Demonstration Kit
Submit Documentation Feedback
Copyright © 2011–2012, Texas Instruments Incorporated