User Manual
7 Maintaining the product
68
7.7.4. Replacing the particle filter
The filter chamber may contain condensate.
1. Open the filter chamber: turn slightly anti-clockwise.
2. Remove the filter cartridge and replace it with a new one
(0554 0040).
3. Attach the filter chamber and lock it: turn slightly clockwise.
Pos: 105 /TD/Produkt ins tand halten/testo 320/ testo 320_Instandh altung Kondensatbehä lter @ 11\mod_13274133 54413_79.docx @ 105353 @ 23 @ 1
7.8. Condensate container
The fill level of the condensate container can be read from the
markings on the condensate trap.
Draining the condensate container
The condensate consists of a weak mix of acids. Avoid skin
contact. Make sure that the condensate does not run over
the housing.
ATTENTION
Damage to the sensors and the flue gas pump due to
condensate entering the gas path!
> Do not empty the condensate container while the flue gas
pump is in operation.