User Manual

7 Maintaining the product
66
1. Loosen and remove halfshell handles (1).
2. Turn the sealing cap (2) clockwise as far as it will go and
remove probe shaft.
3. Release halfshell elements (3) and remove.
4. Remove the adapter (4) and hose (5), push the thermocouple
out of its holder and disconnect the cable (6) from the
thermocouple.
5. Connect cables to the new thermocouple (white -, green +) (7)
and push the thermocouple back into the holder.