User Manual
6 Using the product
57
2. End measurement: [ ].
3. Confirm the message: [OK].
Options:
> [Options] → Clipboard: data is saved to the clipboard.
> [Options] → Delete clipboard: any data saved to the clipboard
is deleted.
> [Options] → Save: the measured values are saved in a record.
> [Options] → Set alarm limits: the alarm limits menu is opened.
Pos: 95 /TD/Produkt ver wenden/testo 320/tes to 320 CO2-Umgebung @ 11\mod_132 7411691882_79.doc x @ 105253 @ 355 @ 1
6.2.15. CO2 ambient
This function is not available when the Spain country version is
selected.
✓ An ambient CO2 probe (0632 1240) must be connected.
In order to obtain correct measured values, it is imperative
to enter the prevailing absolute pressure. This can be
entered directly (Pressure absolute), or it is automatically
calculated when entering Altitude and barometric pressure
(Barometric pressure).
Call up function:
> [ ] → Measurement options → [OK] → AmbCO2 → [OK].
Carrying out the measurement:
1. Select parameter → [Edit].
2. Enter values. [▲], [▼] and in some cases [◄], [►] → [OK].
3. Start measurement: [ ].
4. End measurement: [ ].
- The ambient CO2 value is displayed.
Options:
> [Options] → Clipboard: data is saved to the clipboard.
> [Options] → Delete clipboard: any data saved to the clipboard
is deleted.
> [Options] → Save: the measured values are saved in a record.
> [Options] → Alarm limit: the alarm limits menu is opened.
> [Options] → Edit: values for adjustable parameters can be
edited.
> [Options] → Measurement view: (This function is not available
during a measurement) The measured value display menu is
opened.