User Manual

6 Using the product
52
6.2.7. CO undiluted
Call up function:
A multi-hole probe (0554 5762) should be connected.
> [ ] Measurement options [OK] CO undiluted
[OK].
Carrying out the measurement:
1. Start measurement: [ ]
- The measured value is displayed.
2. End measurement: [ ]
Options:
> [Options] Clipboard: data is saved to the clipboard.
> [Options] Delete clipboard: any data saved to the clipboard
is deleted.
> [Options] Save: the measured values are saved in a record.
Pos: 88 /TD/Produkt ver wenden/testo 320/tes to 320 Rußzahl/WTT @ 13\mod_1 342166204287_79.d ocx @ 137161 @ 355555 @ 1
6.2.8. Smoke number/HCT
Calling up the function:
> [ ] Measurement options [OK] Smoke
number/HCT[OK].
The parameters Smoke No. and Oil depos. are only
available for oil fuels.
Determine smoke tester no./smoke nos./oil depos. with the
smoke pump and enter manually:
1. Select parameter → [Edit].
2. Enter data or values [OK].
Determine smoke tester no./smoke nos./oil depos. with the
smoke tester testo 308 and transmit wirelessly:
- The testo 308 must be in data transfer mode ( lights up).
> [Options] t308.
- The values recorded by the smoke tester are transferred to the
testo 320.