7.
6.2.6. BImSchV
This function is only available when the Germany country version
is selected.
A qA average value measurement can be carried out. For this
purpose the average is determined continuously over a period of 30
s, the measuring cycle takes 1 s. The average values actually valid
at the corresponding time of recording are displayed.
Call up function:
✓ A flue gas probe and a combustion air temperature probe must
be connected.
> [ ] → Measurement options → [OK] → BImSchV → [OK].
> Select the fuel → [OK].
Carrying out the measurement:
1. Start the measurement series: [ ]
Wait for the balancing time, until O
2
shows a value below
20%.
2. [Next].
- The qA measured values (O
2
, AT, VT) are determined (30 s).
- The measurement stops automatically.
- The measured values are displayed and saved automatically in
a record.
3. End measurement: [Close]
or
End measurement and call up draught measurement function:
[Draught - Measuring].
Options:
> [Options] → Clipboard: data is saved to the clipboard.
> [Options] → Delete clipboard: any data saved to the clipboard
is deleted.
> [Options] → Save: the measured values are saved in a record.
> [Options] → Address/Location: the Address/Location folder is
opened.