User Manual

6 Using the product
50
Pos: 85 /TD/Produkt ver wenden/testo 320/tes to 320_Mittelwert (Land esversion IT) @ 13\mod_13 42177433764_79.doc x @ 137329 @ 355 @ 1
5.
6.2.5. Average
This function is only available when the Italy country version is
selected.
Call up function:
A flue gas probe or a multi-hole probe (0554 5762) is
connected.
> [ ] Measurement options [OK] Average [OK].
Options:
> [Options] Recalibrate: the gas sensors are set to zero.
> [Options] Addresse/Location: the Address/Location folder
is opened.
> [Options] Fuels: select fuel.
> [Options] Sensor detection: once the probe has been
changed, start sensor detection manually.
To calculate the average, a series of 3 measurements are carried
out.
Averaging:
1. Position the flue gas probe in the centre of flow (area of the
highest flue gas temperature).
2. Start measurements
> First measurement: [ ].
> Second and third measurement: [OK]
- The set measurement parameters, measurement period and
measured values are displayed.
- A signal is sounded after 2 min (recommended measurement
period)
3. End measurements: [ ].
- Once the series of measurements has been carried out, the
record for averaging is displayed.
> If necessary, scroll through the record: [], [►]
4. [Next]
5. Enter checks:
> Select criterion: [], [].
> Change value: [Edit] [], [] [OK].
6. End check: [Close]
- The record is saved.