Setting limits:
1. Select limit → [Edit].
2. Set values → [OK].
3. Save changes: [Finished].
Pos: 80 /TD/Überschr iften/6.3 Messungen durc hführen @ 0\mod_1184584650 078_79.docx @ 1872 @ 2 @ 1
6.2. Measuring
Pos: 81 /TD/Produkt ver wenden/testo 320/tes to 320 Messung vorbereit en @ 11\mod_1327409352859 _79.docx @ 104989 @ 355554554 44 @ 1
6.2.1. Preparing for measurement
The First steps chapter (see First steps, page 28) must
have been read.
6.2.1.1. Testing for leaks
The entire measurement system (probe, condensate trap, hoses
and connections) must be tested for leaks before each
measurement to avoid incorrect measurements due to the
infiltration of external air. Testing is carried out while the pump is
running and may be performed by attaching a compressed balloon
pump. The measurement system is leak-tight if the balloon pump is
not filled with air.
6.2.1.2. Zeroing phases
Measuring the combustion air temperature
If no combustion air temperature probe is connected, during the
zeroing phase, the measured temperature of the flue gas probe is
taken as the combustion air temperature.
The flue gas probe should not be in the flue gas duct during
the zeroing phase.
All dependent parameters are calculated using this value. This
method of measuring combustion air temperature is sufficient for
systems dependent on ambient air.
If a temperature probe is connected, the combustion air
temperature is measured continuously via this probe.
Gas zeroing
When the instrument is switched on, the measurement menu is
opened and the gas sensors are zeroed.
The flue gas probe must be in fresh air during the zeroing
phase!