Pos: 60 /TD/Produkt ver wenden/testo 330/tes to 330 Rechte Funktions taste belegen @ 6\mod_127928 3093786_79.doc x @ 67168 @ 3 @ 1
6.1.1. Assigning the right function key
The right function key can have a function from the Options menu
assigned to it. The menu Options is accessed via the left function
key and is available in many different menus. This assignment is
only valid for the currently opened menu / the opened function.
✓ A menu / function is opened in which the Options menu is
displayed on the left function key.
1. Press [Options] .
2. Select option: [▲], [▼].
Depending on the menu / function from which the Options menu
was opened, the following functions are available.
3. Assign the selected function to the right function key: Press
[Config. Key].
Pos: 61 /TD/Produkt ver wenden/testo 350 neu/te sto 350 Geräteeinste llungen CU/testo 350 Ger äteeinstellungen_ Ueberschrift @ 5\ mod_1266322627982_79.doc x @ 58663 @ 3 @ 1
6.1.2. Instrument settings
Pos: 62 /TD/Produkt ver wenden/testo 330/tes to 330 Geräteeinstel lungen aufrufen @ 6\mod_1 279008333499_79.doc x @ 65426 @ @ 1
It is assumed that the contents of the chapter First steps
(see First steps, page 28) are known.
Calling up a function:
> [ ] → Device Settings.
see First steps, page 28
Pos: 63 /TD/Produkt ver wenden/testo 320/tes to 320 Messwertanzeige @ 11\ mod_1327406858634 _79.docx @ 104758 @ 4555 @ 1
6.1.2.1. Measurement view
The parameters/units and the display (number of measured values
shown per display page) can be set.
The settings are only valid for the currently chosen measurement
type, which is indicated by the symbol in the info field.
Total overview of selectable measurement parameters and units
(available selection depends on the set country version and
selected measurement type):
Display Measurement parameter
FT
Flue gas temperature
AT
Combustion air temperature